This protocol is to ensure that an accurate patient specific model can be created for planning a dental implant case. Review the following information before proceeding with the scanning process. If you have any questions please contact a ProtoMED representative at 303-466-5610 prior to scanning the patient.
To download a copy of the Dental Implant scanning protocol, right click and save to your desktop.
Area |
FOV |
Spacing |
Slice Thickness |
mA |
Time |
Mandible/Maxilla |
20
cm |
1
mm |
1.5
mm |
120-180 |
2
sec. |
Area |
FOV |
Spacing |
Slice Thickness |
*mA |
Pitch |
Mandible/Maxilla |
20
cm |
N/A |
1
mm |
120-180 |
1 |
The following is a list of CT scanners currently supported at ProtoMED. Please archive image data (not raw data) onto the appropriate media.
FOR BEST RESULTS, PRINT THIS TABLE FROM THE PROTOCOL DOWNLOAD
Scanner Brand |
Image Storage |
Scanner Type |
Media Type |
||||
Pioneer Optical
Disk |
Maxoptix Optical
Disk |
4mm DAT Tape |
8mm DAT Tape |
CD or PTF |
|||
Elscint |
Uncompressed |
CT Twin |
X |
||||
Exel 2400 |
X |
||||||
Twin Flash** |
X |
||||||
GE |
Compressed |
CT PACE |
X |
X |
X |
||
CT/i (DICOM
Images) |
X |
X |
|||||
Hilight Advantage |
X |
x |
|||||
Hispeed Advantage |
X |
X |
X |
||||
Lightspeed Qx/i
(DICOM Images) |
X |
X |
|||||
Prospeed |
X |
X |
X |
||||
Cytec |
X |
X |
X |
||||
Sigma (MR) |
X |
||||||
Hitachi |
Uncompressed |
Hitachi MRP
7000 |
X |
X |
|||
Hitachi W 950 |
X |
X |
|||||
Hitachi W 2000 |
X |
X |
|||||
Phillips |
Uncompressed |
Easy Vision |
X |
X |
|||
Tomoscan CX,
CS/S, CX/Q |
X |
X |
|||||
Tomoscan LX,
LX/C |
X |
X |
|||||
Tomomscan SR |
X |
X |
|||||
Tomoscan 60/TX,
TX |
X |
X |
|||||
Picker |
Uncompressed |
MX Twin Helical
(DICOM Images) |
X |
X |
|||
MX 8000 (DICOM
Images) |
X |
X |
|||||
PQ 2000 |
X |
X |
|||||
PQ 5000 |
X |
X |
|||||
PQ 6000 |
X |
X |
|||||
Siemens |
Uncompressed | Emotion Duo |
X |
X |
|||
Magnetom (MR) |
X |
X |
|||||
Somatom AR |
X |
X |
|||||
Somatom Balance |
X |
X |
|||||
Somatom DRH |
X |
X |
|||||
Somatom HiQ |
X |
X |
|||||
Somatom Plus |
X |
X |
|||||
Somatom Volume
Zoom |
X |
X |
|||||
Toshiba |
Uncompressed |
IS&C Formatted |
X |
X |
|||
If you have a scanner not supported on this list, please call a ProtoMED representative at 303-466-5610 to discuss computability issues.
Siemens® is a registered trademark of Siemens AG. General Electric® is a registered trademark of General Electric Company. Picker® is a registered trademark of Picker International, Inc. Toshiba® is a registered trademark of Toshiba Corporation. Elscint® is a registered trademark of Elscint, Inc.